ANI Monitor V1 - Continouous analgesia monitoring system
MD/PRD/IN16.ANIV1 V.10 – 29 SEP 2021
11
Do not block ventilation inlet holes on the sides of monitor.
Using accessories other than those specified may result in increased electromagnetic
emissions or decreased electromagnetic immunity of the ANI Monitor V1.
Using electrode arrays different from those sold by Mdoloris Medical Systems could damage
the device or lead a risk for the user or the patient.
Use again an array could reduce the adhesion, leading to a possible decrease in electrical
performance.
Use again an array could damage the capacitor due to application, removal and reapplication.
Only Mdoloris Medical Systems qualified biomedical technicians must perform service or
repairs on ANI Monitor V1.
The ANI Monitor V1 contains an internal battery. This battery must be removed by an
authorized technician and discarded or recycled in accordance with national regulatory in
force in the country. Contact Mdoloris Medical Systems or the local distributor for the battery
maintenance.
The internal battery is not designed to withstand alone a long time monitoring. The internal
battery of the ANI Monitor V1 allows the system to keep calculating the ANI while the
patient is moved to another room or in case of temporary interruption of electrical main
supply.
To ignite the monitor, it has to be plugged in first intention to prevent a sudden electrical
crash in case of the battery is not enough charged. Check the battery charge symbol displayed
before unplugged: the battery could be not sufficiently charged and interrupt the electrical
supply of the ANI Monitor V1.