Programming instructions 

Software Version 5 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0250032E/ February 2014 

© MAXX GmbH, 2014. All rights reserved. Printed in Germany 

Содержание SP-TP-P6

Страница 1: ...Programming instructions Software Version 5 0250032E February 2014 MAXX GmbH 2014 All rights reserved Printed in Germany...

Страница 2: ...th explanation 10 Examples of programming 14 Programming of a time proportional sampling program 14 SPECIAL PROGRAM FUNCTIONS 16 SPECIAL FUNCTIONS 16 PROGRAMMING OK 16 SERIAL SAMPLES 16 BOTTLE ASSIGNM...

Страница 3: ...H SAMPLING 25 BATCH SAMPLING 25 FREELY PROGRAMMABLE INPUTS 26 OUTPUT SIGNALS 27 MESSAGES description 28 ERROR MESSAGES 28 LOG MESSAGES MEMORY 30 ANNEX VAR variable sampling system 32 CALIBRATION 32 AN...

Страница 4: ...Confirm the choice is automatically marked Entry change of values Confirmation of entered values Return to higher menu level Initialization reset display Terminate sleep mode only portable RESET rese...

Страница 5: ...the display indicates that there are further selection possibilities Menu variants The top line indicates that you can navigate to the right or to the left by pressing the arrow keys The bottom line...

Страница 6: ...n the right side of the window indicates that there are further entries which can be displayed by scrolling up or down Depending on the menu the display will show in which direction you can scroll Sel...

Страница 7: ...age 7 MAIN MENU STRUCTURE Description of the main menu structure with submenu levels 2 and 3 PROGRAMS STATUS STOP MANUAL SAMPLING DATA MEMORY INFO PAUSE STOP IMMEDIATELY DATE TIME WEEKDAY TIME TIME FL...

Страница 8: ...ION INFO IP ADDRESS OUTPUTS TEMPCARD PUMP PINCH VALVE VALVE SYSTEM DISTRIBUTOR DIGITAL OUTPUTS FLOW DIGITAL EVENT DI3 DI4 DI5 DI6 DI7 DI8 ANALOG 1 ANALOG 2 ELECTRODES 1 ELECTRODES 2 PT 1000 SENSOR OP...

Страница 9: ...NGS DD MM YYYY hh mm 15 08 2013 13 56 LANGUAGE DISTRIBUTOR MAX SUCTION TIME PRE PURGE POST PURGE CALIBRATE VAR SYSTEM AERATION TIME PUMP POWER INTERNAL TEMP ANALOG SIGNAL DISPLAY PAUSE DURATION FREELY...

Страница 10: ...m start can be immediately with date time dd mm yyyy hh mm WEEKDAY TIME with weekday time day hh mm CHANGE PROGRAM No xx Change the program parameters like mode of operation time flow event interval e...

Страница 11: ...PCARD Display STATUS of Lower heating OFF ON Cooling OFF ON Upper heating OFF ON DIGITAL INPUTS Display of DI digital input Flow digital 0 Event 0 DI3 DI4 DI5 DI6 DI7 DI8 ANALOG INPUTS Display of ANAL...

Страница 12: ...tion AERATION TIME Time until pinch valve is opened for drain off of sample PUMP POWER Adjustable from 70 to 100 INTERNAL TEMPERATURE Via NTC Via PT1000 Limit value 1 20 C Delay time 1 60 min Example...

Страница 13: ...E Only at portable samplers ACTIVE If the sleep mode has been activated and the program is to be started in 20 min at the earliest the message Attention device switches to sleep mode is displayed for...

Страница 14: ...ng program Select PROGRAMS in the main menu Select CHANGE Select PROGRAM NO 1 out of 12 Programs No 2 12 can be selected by pressing the left or right arrow key Press ENTER to edit the program Selecti...

Страница 15: ...02 up to 168 00 hhh mm Programming can be terminated and the program can be started directly Besides the standard programming there are several special program functions which can be activated select...

Страница 16: ...e value 3 is entered 3 samples are extracted successively When activating this function particular attention has to be paid to the bottle volume in order to avoid overfilling This function is useful i...

Страница 17: ...and leads to a deferred start by the time entered of the next program Example program 1 is edited with a program pause of 1 hour and is started at 8 00 h 24 hour cycle Thus the program will be termin...

Страница 18: ...hh mm Program run As soon as there is an event signal the distributor moves to the next empty bottle is recorded in the memory as event bottle The sample extraction is effected according to the set va...

Страница 19: ...lse is before or after the programmed start time here 8 00 h External pulse is triggered before the end of the program run time Example you would like to change the bottles already before the expiry o...

Страница 20: ...ed bottle filling pause A bottle change is made every 2 hours that means after 24 hours all the 12 bottles are filled with programmed bottle filling pause of 24 hours 1440 min Bottle 1 is filled for 2...

Страница 21: ...he first requested sample extraction The program start is effected on the selected weekday and at the selected time in the _____________________________________________________________________________...

Страница 22: ...S STOP Here the status active inactive of programs is displayed The status of programs 1 12 can be checked by pressing the arrow keys right left STATUS STOP When pressing the Enter key the following i...

Страница 23: ...menu SETTINGS The pause can be terminated manually or it is automatically terminated after the entered xxx An active program can be stopped aborted If several programs are active all these programs t...

Страница 24: ...t the sampler to the plant s signal Set 0 4 and 20 mA by means of the SPS or an analog signal transmitter and confirm these values according to the menu instructions 1 Connect 0 4 mA and confirm 2 Con...

Страница 25: ...ow dependent analog and digital The programming is effected as described before sampling mode time flow digital BATCH SAMPLING Other than at the classic event sampling the bottle change is not effecte...

Страница 26: ...LSE if selected program x will be stopped terminate with ESC PROGRAM RUN DURING PULSE A program is executed as long as there is a continuous signal If the signal drops the program will be stopped BOTT...

Страница 27: ...SIGNAL 2 OUTPUT SIGNAL 3 OUTPUT SIGNAL 4 OUTPUT SIGNAL 5 Each output signal 1 5 can be programmed individually according to the following list PROGRAM ACTIVE Selection PROGRAM ACTIVE or PROGRAM XX ACT...

Страница 28: ...F Deactivation of output signal ERROR MESSAGES Error code Text meaning Description 1 ERROR DISTRIBUTOR Distributor is blocked pulse generator or light barrier are defective 2 ERROR SUCTION No water ho...

Страница 29: ...in VAR or pneumatic operation 15 EMERGENCY CUTOFF Current flow at an output of the controller is too high or there is a short circuit 1 error at a digital output 2 pinch valve valve system error 3 ov...

Страница 30: ...ge triggered via a programmable input only possible if no program is active 14 VOLTAGE LOSS START Date time of the start of a power failure 15 EVENT START Date time of the start of an event 16 END OF...

Страница 31: ...r limit value 29 ANALOG VALUE X mA signal values Logging interval each xx minutes 1st value No of channel 2nd value average value of logging interval 3rd value upper limit value has to be ordered sepa...

Страница 32: ...extracted Thus the sample extraction is always proportional to the flow At this system the sampling interval is fixed e g every 10 min and the sample volume varies the volume changes according to the...

Страница 33: ...hich are started automatically Please also collect the samples of all 3 cycles in a measuring glass 6 ACTUAL VALUE upper value Here the total volume of all 3 cycles of the upper value calibration is e...

Страница 34: ...23 If a DHCP server is not available the default IP is 192 168 60 8 2 Check the IP address at the sampler In case of an operation in a local network the IP address is received via DHCP or can be perma...

Страница 35: ...in connect 1 Click on the file connect exe to start the installation 2 If during the installation process a security remark regarding the driver software is displayed click on Install this driver any...