Page 31
23
ACCESS WITH PASSWORD
Date/time of access to a menu which requires the
extended password, e.g. service menu, stop
program, change settings etc.
24
PT1000 °C/U-BATT
Temperature values of the PT1000 sensors as
well as the operating voltage of the controller.
Logging interval: every 10 minutes.
26
SPL REQUESTED
(REMOTE)
Sample extraction requested via a programmable
input (only possible if no program is active)
27
SPL REQUESTED (EVENT)
Sample extraction triggered via an event program
28
ANALOG VALUE A1
mA signal values, logging interval: every 10
minutes
(current value, no average value
calculation)
1st value: measured value at logging time
2nd value: lower limit value
3rd value: upper limit value
29
ANALOG VALUE X
mA signal values. Logging interval: each xx
minutes.
1st value: No. of channel
2nd value: average value of logging interval
3rd value: upper limit value
(has to be ordered separately by the customer)
30
OVERFILL PROTECTION
Value
1
= Drop sample
2
= Switch to next bottle
The overfill protection function has been activated
in flow-dependent sampling mode
1st value: the requested sample was dropped.
2nd value: the requested sample was filled into the
next bottle.
35
TOTAL VOLUME
At the program stop the total volume of all samples
requested during the program run is added up
(only at VAR and peristaltic pump systems)
36
FLOW WHEN SAMPLING
Logging of flow at the time a sample extraction has
been triggered (only at VAR and peristaltic pump
systems)
37
RWA-DWA STATUS
Values are logged when the rain weather pulse
divider is activated or deactivated. Selection is
made per weekday from 00:00 – 23:59. (Only
available at the Limburg sampler version)