3.
ACTUAL VALUE lower value
Please enter the total volume of the 3 cycles.
4.
INFO after the completion of the calibration of the lower value.
The displayed value has no relation to the volume.
5.
Start
upper
value
Again there are 3 consecutive calibration cycles which are started
automatically.
Please also collect the samples of all 3 cycles in a measuring glass!
6.
ACTUAL VALUE upper value
Here the total volume of all 3 cycles of the upper value calibration is
entered.
7.
INFO after the completion of the calibration of the upper value
This value corresponds to the average volume in ml.
Now, the system is calibrated and can be operated.
REMARK
:
A correct functioning is only possible when the analog input has been
synchronized/calibrated to the signal of the plant! Otherwise there might be wrong sample
volumes!