5-5
Advanced Display Modes
Defining a Unit Label
To define a unit label, do the following:
1.
Ensure that a BetaPort–DC module is installed in
one of the module bays, then press
CH1
or
CH2
to
indicate the appropriate bay in which the DC
module is installed.
2.
If necessary, use the
V/mA
Setup menu item to
change the electrical units to match the field
instrument's electrical signal.
3.
Use the
Menu
key to locate the Setup menu item
User
, then press the corresponding function key to
display the following screen:
In the above example:
x
= the channel number of the DC module,
yyyy
= the current user unit label.
4.
Specify a unit label by using the
CLR
key to erase
the character (if any) to the left of the cursor and
use the arrow keys on the keypad to scroll through
the available character set. When the desired
character is displayed, press the
ENTER
key to
accept the character and move the cursor to the
right. A maximum of 4 characters may be entered
for identification.
NOTE:
Pressing the
ENTER
key twice in a row
ends editing. Therefore, to specify the same
character consecutively, use the up and down arrow
keys to select the second character.
5.
When you have completed entering the unit label,
press the
ENTER
key once again to return to the
Setup menu.
CHx Ent. Unit: yyyy
Содержание BetaGauge II
Страница 1: ...BetaGauge II Pressure Calibrator User Guide ...
Страница 8: ...viii Specifying Post Test Status Information 7 23 ...
Страница 11: ...xi ...
Страница 15: ...xiv ...
Страница 27: ...General Description 1 12 ...
Страница 35: ...Input Modules 2 8 ...
Страница 141: ...Application Notes 6 56 ...
Страница 165: ...Doc Cal Mode 7 24 ...
Страница 183: ...Specifications B 8 ...
Страница 189: ...Service Information D 4 ...
Страница 197: ...Index I 8 ...