![Martel BetaGauge II Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/martel/betagauge-ii/betagauge-ii_user-manual_1732374049.webp)
Basic Operations
3-14
3.
Press the function key corresponding to the
damping setting you want to use, then press the
CLR
key to return to the Setup menu.
CAUTION:
Enabling the damping feature affects
your ability to accurately zero a module’s
displayed value. Therefore, prior to
zeroing the displayed value of a module,
disable the damping feature by setting the
damping value to 0.0.
Viewing Module Information
The BetaGauge II allows you to view the following
information for the modules currently installed in the
calibrator:
Serial number
Full scale range (pressure modules only)
Last calibration date
Procedure
To view module information, do the following:
1.
Use the
Menu
key to locate the Setup menu item
Mod
, then press the corresponding function key to
display the Module Info menu.
2.
Press the function key corresponding to the channel
you want to view. Remember that
CH1
represents
the left module and
CH2
represents the right
module. The lower portion of the LCD will appear
similar to the examples shown below:
DC Module Example:
DC SN: 610006
Cal: 09/25/2005
Содержание BetaGauge II
Страница 1: ...BetaGauge II Pressure Calibrator User Guide ...
Страница 8: ...viii Specifying Post Test Status Information 7 23 ...
Страница 11: ...xi ...
Страница 15: ...xiv ...
Страница 27: ...General Description 1 12 ...
Страница 35: ...Input Modules 2 8 ...
Страница 141: ...Application Notes 6 56 ...
Страница 165: ...Doc Cal Mode 7 24 ...
Страница 183: ...Specifications B 8 ...
Страница 189: ...Service Information D 4 ...
Страница 197: ...Index I 8 ...