![Martel BetaGauge II Скачать руководство пользователя страница 181](http://html1.mh-extra.com/html/martel/betagauge-ii/betagauge-ii_user-manual_1732374181.webp)
Specifications
B-6
Environmental
Temperature
-20ºC to 50ºC (operating)
0ºC to 35ºC (storage with battery packs)
-30ºC to 70ºC (storage without battery packs)
Relative Humidity
0 to 95% non-condensing
Effects
Pressure:
± 0.0015% Range/ºC outside 15ºC to 35ºC
Electrical:
±0.002% Range/ ºC outside 15ºC to 35ºC
Overpressure
Non-isolated Sensors
0 to 5 psi
400% of sensor range
>5 psi
300% of sensor range
150 psi
200% of sensor range
Isolated Sensors
200% of sensor range
Alarm Trip Detect
Wet contact input
12 Vdc ± 10% threshold, 48 Vdc maximum
Dry contact input
800 Ohm ± 200 Ohm threshold,
4 mA maximum excitation current
Damping Factor
Programmable 2, 5, or 10 seconds response time for all
inputs (running average)