7-23
Doc. Cal. Mode
Specifying Post-Test Status Information
NOTE:
You must answer the post-test status prompts.
If you press the
CLR
key prior to answering
the post-test prompts, all calibration results
will be lost.
To specify post-test status information, do the
following;
1.
From the Post-Test Status screen, use the function
button corresponding to the appropriate menu item
(
Yes
,
No
, or
Rep
) to indicate whether the field
instrument was adjusted, left alone, or replaced.
2.
After you have specified the field instrument
adjustment status information, the BetaGauge II
prompts you for an identification number of the
technician who performed the calibration test.
3.
Do one of the following:
Use the numbers on the keypad to specify a
technician’s identification number, then press
the
ENTER
key.
Accept the 0 default value and press the
ENTER
key to bypass the entry.
4.
After you have specified the identification number,
or bypassed the prompt, the display returns to the
Tag Status screen of the Main level. Do one of the
following:
Press the
CLR
key to exit the SYS mode and
return to the Setup menu.
View the results from the last calibration
performed on the selected tag. (Refer to the
section entitled
View the Calibration Results
for a Specific Tag
earlier in this section.)
Select another tag.
Содержание BetaGauge II
Страница 1: ...BetaGauge II Pressure Calibrator User Guide ...
Страница 8: ...viii Specifying Post Test Status Information 7 23 ...
Страница 11: ...xi ...
Страница 15: ...xiv ...
Страница 27: ...General Description 1 12 ...
Страница 35: ...Input Modules 2 8 ...
Страница 141: ...Application Notes 6 56 ...
Страница 165: ...Doc Cal Mode 7 24 ...
Страница 183: ...Specifications B 8 ...
Страница 189: ...Service Information D 4 ...
Страница 197: ...Index I 8 ...