Basic Operations
3-4
BetaPort–P Pressure Module
The BetaPort– P pressure module can display in any of
the following units:
PSI
InHg
InWC
mmHg
cmWC
kPa
bar
mbar
atm
kpcm
MPa
(User)
Upon initialization, the BetaPort–P pressure module
defaults to PSI pressure units. Use the following
procedure to select other pre-programmed pressure
units:
1.
Press
CH 1
or
CH 2
, as appropriate, to select the
BetaPort–P pressure module for modification.
2.
Press the function key corresponding to
Unit
to
view the list of pre-programmed units
.
The
menu item allows you to scroll the menu for other
selections.
NOTE:
If you select
Unit
for a channel where the
BetaPort–DC module is installed, the message
Invalid Selection
will appear. Press the function
key corresponding to
Clear
to erase this message
and return to the Setup menu.
3.
When the pressure unit you want to use appears,
press the corresponding function key to select the
unit. You can view the current selection in the LCD
display for the appropriate channel.
NOTE:
If you select
InWC
or
cmWC
, refer to the
section
Water Column Measurements
on the
following page.
Содержание BetaGauge II
Страница 1: ...BetaGauge II Pressure Calibrator User Guide ...
Страница 8: ...viii Specifying Post Test Status Information 7 23 ...
Страница 11: ...xi ...
Страница 15: ...xiv ...
Страница 27: ...General Description 1 12 ...
Страница 35: ...Input Modules 2 8 ...
Страница 141: ...Application Notes 6 56 ...
Страница 165: ...Doc Cal Mode 7 24 ...
Страница 183: ...Specifications B 8 ...
Страница 189: ...Service Information D 4 ...
Страница 197: ...Index I 8 ...