Application Notes
6-28
Pressure to Pressure Test
You can perform a pressure to pressure calibration test
using either the conventional mode or the SYS mode of
operation. The following paragraphs describe all of the
possible configurations, as well as, the other equipment
necessary to perform a pressure to pressure test.
Conventional Mode
You may perform this test if:
Your field instrument has a pressure input
range within the ability of one pressure module
on the BetaGauge II, and
Your field instrument has a pressure output
range within the ability of the other pressure
module on the BetaGauge II.
Field Instrument Setup
Pressure input
Pressure output
Other Necessary Equipment
Pressure source
Содержание BetaGauge II