background image

Manu

Flo

 

 

 

 

 

       

Page 5

 

 

             

MANU ELECTRONICS

 

PTY LTD

 

Flow Measurement Products

                                                                                            

41 Carter Road, Brookvale NSW 2100 

                                        

                         

 

                                   

Website: 

http://www.manuelectronics.com.au

 

 

RPFS_Datasheet 23/02/2018

 

 

ADMIX

STORAGE

TANK

Filter

Non-Return

Valve

tank

Valve

outlet line 
above level of
Storage Tank
(most common)

Gate Valve 
restricts flow

Flowmeter

Non-Return Valve

optional  3-Way Valve 
(calibration point)

optional  240vac 

Solenoid

to 

prevent  Venturi effect when 
injecting into water lines.

VAC

SUPPLY

E  N  A

A

N

TWIN 
SHIELDED 
CABLE

at rear of 

Batch 

Controller

PUMP

ME995

WIRING

DIAGRAM

ACTIVE

Optional batch override/top-up 
momentary push button switch

 

Standard AC Wiring for Pump and optional Solenoid      

 

                                                                                                

at rear of 

Batch 

Controller

ACTIVE

Rear

ME995

SHIELD (0 v)

PULSE

SHIELDED

CABLE

CONTACT

CONTACT 1

+VDC

Optional batch override/top‐up 
momentary push button switch

Gate Valve

Solenoid

Flowmeter

‐‐

+

+VDC

0v

Open Contact
Output Drive

12 – 24 

VDC supply

0v

+VDC

EARTH 
(GROUND)

liquid

+ 12 VDC

0v 

+

PUMP

 

 

Wiring for DC-powered Batch Controller with DC Open Contact Output Drive to Pump and/or Solenoid 

NOTE: if current draw of solenoid is > 0.5 Amps, or if using a pump, then install a contactor 

 
 
 
 

ME995-7

 

Содержание ManuFlo ME995-7

Страница 1: ...urrent draw Consider wiring an override button N O with spring return for manual batching or top up of water which will be counted by the controller display See wiring diagram C Install the Rota Pulse paddlewheel sensor RPFS P in a straight pipe section with the same diameter pipe as the adaptor tee section with 10x diameter before and 5x diameter after the sensor with no elbows reducers valves or...

Страница 2: ...1 Litre increments up to a maximum 9000 Litres The controller operates from standard 220 260 vac or optional 24vac 110vac or 12 24 VDC voltage supplies Contact output drive is via one or optional two relays or with OC open contact drive to allow control output switching drive of any external voltage Standard controllers are in panel mount form or optionally can be housed in a metal box or IP65 ABS...

Страница 3: ...ected 200L when 190L on LEDs this is 10L more or 5 over 10 190x100 So decrease the calibration value by 5 i e if calibration set to 300 new value is 300 5 300 15 285 Set Calibration U 5 T 8 H 2 4 If the amount collected is less than is shown on the LED display then increase the set calibration value by the same difference e g if collected 180L when 190L on LEDs this is 10L less or 5 under 10 190x1...

Страница 4: ...wo product changeover output drive Allows 2 flowmeter inputs pump drives e g ME995 7 is the standard Batch Controller 240vac powered whereas ME995 7 MC2 is an ME995 7 Batch Controller with a scaled open collector pulse output and a Start Stop Reset PLC commands interface HOUSING ENCLOSURES SHB SHB1 Single enclosure Powder coated metal Single enclosure Powder coated metal Wired with 240vac contacto...

Страница 5: ... N TWIN SHIELDED CABLE at rear of Batch Controller PUMP ME995 WIRING DIAGRAM ACTIVE Optional batch override top up momentary push button switch Standard AC Wiring for Pump and optional Solenoid at rear of Batch Controller ACTIVE Rear ME995 SHIELD 0 v PULSE SHIELDED CABLE CONTACT CONTACT 1 VDC Optionalbatch override top up momentarypush button switch Gate Valve Solenoid Flowmeter VDC 0v Open Contac...

Страница 6: ...tion to larger size pipes is possible depending on pipe wall thickness alternatively the Long Stem LS versions with adaptors are then used With only one moving part and limited intrusion into the pipe and combined with its flow through design the RPFS allows accurate measurement of liquid flows with virtually no head losses Each of the 4 blades of the rotor paddlewheel extends approximately one ce...

Страница 7: ...val of RPFS from Pipe adaptor Fitting Triangular Keyway Type Nipple Adaptor see FIG 6 1 Unscrew the black PVC locking cap anti clockwise 2 Hold the neck of the Tee piece in your left hand grasp the RPFS body with your right hand and turn slowly anti clockwise until the sensor hydraulically raises out of slot then pull upwards out of the socket never by the cable When returning the sensor to nipple...

Страница 8: ...talled in a horizontal inclined or vertical pipe position Note If mounted in horizontal or inclined pipe make sure insertion position of sensor is at top or 45 from top not on the underside Selection of pipe diameter For best results use the table below Pipe size Flowrange Litres min Pulses Litre mm Min Max approx 1 2 20 13 160 116 25 20 250 75 32 30 410 46 40 50 640 30 50 90 1000 20 63 132 1580 1...

Страница 9: ... SC140 LS 150 mm PVC150SC LS SCP150 LS SC160 LS 160 mm SCPE160 LS SC160 LS 200 mm PVC200SC LS SCP200 LS SC200 LS 225 mm SCPE225 LS SC225 LS 250 mm SCP250 LS SC250 LS 280 mm SC280 LS 300 mm PVC300SC LS SCPE300 LS SC315 LS 315 mm SC315 LS 500 mm Galvanised iron threaded ends BSP female 2000 kPa Note 25mm can be supplied with straight pipe sections already fitted Part GAL25 T2 PVC T piece Class 18 Ca...

Страница 10: ...RPFS and ME995 Wiring and Plumbing Diagrams c 2004 Manu Electronics Pty Ltd 1 ...

Страница 11: ...RPFS and ME995 Wiring and Plumbing Diagrams c 2004 Manu Electronics Pty Ltd 2 ...

Страница 12: ...RPFS and ME995 Wiring and Plumbing Diagrams c 2004 Manu Electronics Pty Ltd 3 ...

Страница 13: ... past batch select Faulty solenoid valve not closing service solenoid valve check air pressure or water does not stop properly insufficient air pressure Wet loads more water Paddlewheel bushes worn Check paddlewheel collected than indicated Wheel dirty flowing over range Replace with new paddlewheel recalibrate Dry loads less water Requires recalibration test Set new calibration figure rear switch...

Страница 14: ...c minimum flowrate required If flow drops below 1 5 Litres sec the pulse fail will activate The equation is Pulse fail frequency Hz Pulses Litre x minimum flowrate of pipe diameter in Litres sec The flowrate frequency Hz minimum setting can be adjusted by soldering a capacitor in parallel with the standard capacitor found on the PCB See Table 2 and diagram below for values and location on the PCB ...