43
58-641 PULSAR Model R86 Pulse Burst Radar Transmitter - F
OUNDATION
fieldbus
™
Instruction Manual
Home Screen
Main Menu
Device Setup
Identity
Basic Config
Display Config
Advanced Config
Factory Config
Sensitivity:
1 to 255
Top Blocking Distance:
12 to 120 inches
(30 cm to 3 meters)
Bottom Blocking Distance:
0 to 120 inches
(0 to 3 meters)
SAFETY ZONE SETTINGS
Safety Zone Alarm:
Off
On
Latch
Safety Zone Height:
(not used when Safety Alarm is
Off)
2 inches to 20 feet
(5 cm to 6 meters)
Reset SZ Alarm
(used when Safety Alarm is
Latch)
Echo Loss Delay:
1 to 1000 seconds
Failure Alarm Delay:
0 to 5 seconds
Level Trim:
-24 to +24 inches
(-60 to +60 cm)
THRESHOLD SETTINGS
Target Selection:
First Echo
Largest Echo
First Moving Echo
Target Thresh Mode:
Automatic
Fixed Value
Target Thresh Value:
0-99% (Auto Mode)
0–253 (Fixed Mode)
Base Threshold:
0–99 ESU
TIME VARIABLE GAIN:
TVG Start Value
TVG Start Location
TVG End Value
TVG End Location
# Run Average
Max Surface Velocity
Max Level Jump
Empty State Delay
Compound Peak Logic
New User Password:
0 to 59,999
CONFIG CHANGED:
Indicator Mode:
Disabled
Enabled
Reset Config Chngd:
Reset?
No
Yes
Reset Parameters:
No
Yes
Home Screen
Main Menu
Device Setup
Identity
Basic Config
Display Config
Advanced Config
Factory Config
NAP
Factory Reset
FACTORY CALIBRATION
(Factory password required)
Elec Temp Offset
Conversion Factor
Scale Offset
Fiducial Gain
Fiducial Strength (read only)
Initial Gain
TVG Divisor
4.9
Model R86 Configuration Menu — Advanced/Factory Configuration