190
47DT OPERATING MANUAL EM-9508-B Rev.7
27.2 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the settings.
1
Hold down Alarm/
↓
, Init/
↑
and Shift buttons at once for 5 sec-
onds or more to move on to Loop Test Output Mode.
• The measuring is stopped and the last measured values or status
are held for the DC, alarm and BCD outputs.
• The current indication starts blinking, to which you can apply
changes.
• The SD indicates ‘TEST UP’.
NOTE
• In moving on to Loop Test Output Mode while ‘S.ERR’ is indicated, min. value of the usable range is indicated when the
SD indicates ‘UNDER’, and max. value is indicated when the SD indicates ‘OVER’.
• In moving on to Loop Test Output Mode while ‘B.ERR’ is indicated, max. value of the usable range is indicated when the
burnout is set to upscale, and min. value is indicated when the burnout is set to downscale.
• The specified decimal point position is applied.
1,3
1,3
1,2
2
Blinking