137
47DT OPERATING MANUAL EM-9508-B Rev.7
16.1 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the specifications and settings.
1
Hold down Alarm/
↓
and Init/
↑
buttons at once for 3 seconds or
more to move on to Advanced Setting Mode.
• The averaging type is indicated.
• The SD indicates ‘AVE-TP’.
NOTE
With the I/O option code ‘6’ or ‘A’, the event trigger mode is indicated (‘EVENT’ on the SD).
2
Press Alarm/
↓
or Init/
↑
button to go to the automatic return
time to Measuring Mode setting.
• The automatic return time to Measuring Mode is indicated.
• The SD indicates ‘RETURN’.
3
Press Shift button to shift the display into the setting standby
mode. Then press Shift and Up buttons to set the automatic
return time to Measuring Mode.
• The second digit starts blinking, to which you can apply changes.
• Set within the range of 00 to 99.
1,2,
4,5
1,2,
4,5
3
3
Blinking