Maintenance
48
BA-42820-02-V03
Water sampling station
EASYPRO
Operating instructions
Reset to the factory settings
You can now reset the device to its factory settings. This deletes the con-
figuration. You must then either load a configuration file or perform the
configuration manually.
Perform the following work steps:
1.
Working in the main menu, navigate to System > Service > Device.
2.
Press “factory settings”.
3.
Confirm with “Yes”.
4
The configuration will be deleted. You must proceed with the follow-
ing section.
ü
All factory default settings will be reset.
Reset the configuration
The device configuration will be saved in *.SET files. A factory-set config-
uration file with standard settings is already present. You can change
these or save your personal configuration in new files.
Recommendation:
Leave the factory-set configuration file unchanged
and save your personal configuration in a new file. Given problems with
the configuration, this enables you to return to a functioning configuration
quickly.
Perform the following work steps:
1.
Working in the main menu, navigate to System > Settings > Configu-
ration.
2.
Select an existing configuration file.
3.
Click “Load” to confirm.
4
The device configuration returns to the saved state.
ü
Load the old configuration.
11.11 Finishing maintenance
Perform the following work steps:
1.
Make a note of the date and scope of the maintenance performed.
2.
In the “Service” menu, navigate to the “Service entry” tab. Enter your
company name and notes about the maintenance. Activate the re-
minder function and enter a date for the next service. Confirm with
the “Save” button .
4
Your service action has been saved in the logfiles.
3.
To restart the system, proceed in accordance with the instructions in
section 8 „Commissioning of the water sampling station“ on page
28.
ü
Maintenance completed.