Configuration menu
27
© Lutz-Jesco GmbH 2020
Subject to technical changes.
200922
BA-62010-01-V01
Chlorine Dioxide System
EASYZON Dd
Operating Instructions
è
Press
up
or
down
to scroll the various parameters; press
Enter
to
enter the
Edit
mode, press
up
or
down
to edit the field or the
description, then press
Enter
again to confirm.
Configuration outputs 10, 11, 12
(not for compact models)
It is possible to set the digital outputs (10, 11 and 12 relays), couple them
to one of the following parameters or to the operation mode:
n
Min level reagent alert
(with switch level linked)
n
Pre-alarm gas detector
(with probe linked)
n
Min ClO
2
Alert
(with probe linked)
n
Max ClO
2
Alert
(with probe linked)
n
Max Chlorite Alert
(with probe linked)
n
Extra max batch lev. Alert
(only in batch mode)
n
Batch Load Solenoid Com.
(only in batch mode)
n
Lack sample flow alarm
(with probe linked)
è
Press
up
or
down
to scroll the various parameters; press
Enter
to
enter the
Edit
mode, press
up
or
down
to edit the field or the
description, then press
Enter
again to confirm. Repeat the
operations above to configure all the outputs.
Configuration inputs 8, 9, 10, 11, 12
It is possible to set the inputs configuration 8 – 12 (8, 9 and 12 already
have default settings), couple them to one of the following parameters or
to the operation mode:
n
Acid min level switch
(input where link the probe)
n
Chlorite min level switch
(input where link the probe)
n
Manual mode reduct. (%)
n
Analys. Sample flow sensor
(input where link the probe)
n
Batch Storage min. Lev.
(only in batch mode)
n
Batch Storage max. Lev.
(only in batch mode)
n
Batch Storage ex.max. lev.
(only in batch mode)
è
Press
up
or
down
to scroll the various parameters; press
Enter
to
enter the
Edit
mode, press
up
or
down
to edit the field or the
description, then press
Enter
again to confirm. Repeat the
operations above to configure all the outputs.
9.4.1.3 Probes Parameters
This menu allows the operator to enable the analysis probe for gas and
chemicals level taks. Furthermore it is possible to set the parameters of
the probes.
n
Residual Chlorine Dioxide
n
Chlorite
n
Gas detector
n
Reagents levels measuring
è
Press
up
or
down
to scroll the various parameters; press
Enter
to
Enter the
Edit
mode, press
up
or
down
to edit the field or the
description, then press
Enter
again to confirm.
Residual ClO
2
sensor
(not for compact models)
After the connection between the sensor (or the output of an analysis ins-
trument with “extern” sensor) and the AI2 input (input 4 – 20 mA, termi-
nals 77 – 78 on the junction box) the operator shall set the parameters as
follow.
n
Sensor
: Absent
→
Through this parameter it is possibile to confirm
the presence of the sensor.
n
AI Sensor Input
: --
→
Set the analogic input number linked to the the
sensor. Actually the only available input is the 02.
n
F.Scale. an. probe (mg/l)
: 0.5
→
Setting this parameter the operator
can insert the sensor full – scale value and then the corrispondent va-
lue at 20 mA.
n
Minimum level (mg/l)
: 0.00
→
Trough this parameter it is possibile
to set a threshold in order to signal a minimum level. It is possible to
couple it with a relay output for remote control.
n
Maximum level (mg/l)
: 0.50
→
Trough this parameter it is possibile
to set a threshold in order to signal a maximum level. It is possible to
couple it with a relay output for remote control.
n
Max production off
: NO
→
Switching on “YES” the operator will ob-
tain the production stop after passing the maximum threshold seen in
the previous paragraph. The production will start again after the misu-
ared value will be lower than the threshold minus the hysteresis.
n
Hysteresis (mg/l)
: 0.01
→
This is set on the hysteresis value concer-
ning the maximum and the minimum level.
n
Alarm delay (s)
: 060
→
Set the alarm delay compared to the warning
for passing both threshold.
For the first calibration and following corrections see chapter 9.3.3 „Cali-
bration of the dioxide sensor“ on page 24.
If the the probe is a passive one, in other words it needs external power
supply, the operator shall switch the jumper cable as in the follwing pic-
tures.
è
Press
up
or
down
to scroll the various parameters; press
Enter
to
Enter the
Edit
mode, press
up
or
down
to edit the field or the
description, then press
Enter
again to confirm. Press
BACK
and
up
or
down
to move to another sensor.
Chlorite Sensor
(not for compact models)
After the connection between the sensor (or the output of an analysis ins-
truments with “extern” sensor) and the AI4 input (input 4 – 20 mA, termi-
nals 79 – 80 on the junction box) the operator shall set the parameters as
follow.
n
Sensor
: Absent
→
Through this parameter it is possibile t confirm the
presence of the sensor.
n
AI Sensor input
: --
→
Set the analogic input number linked to the the
sensor. Actually the only available input is the 04.
n
F.Scale. an probe (mg/l)
: 2.0
→
Setting this parameter the operator
can insert the sensor full – scale value and then the corrispondent va-
lue at 20 mA.
n
Maximum level (mg/l)
: 0.80
→
Trough this parameter it is possibile
to set a threshold in order to signal a maximum level. It is possible to
couple it with a relay output for remote control.
n
Hysteresis (mg/l)
: 0.01
→
This is set on the hysteris value concer-
ning the maximum and the minimum level.
n
Alarm delay (s)
: 060
→
Set the alarm delay compared to the warning
for passing both threshold.