Symptom
Possible Causes
Recommended Solutions
Event rate is
slower than
expected
Sample concentration
is low
Make sure the sample concentration is between 10
7
and 10
8
cells/ml. Lower
concentrations can be used but this will decrease the cells/second.
Core is off center
Cropped images will be eliminated from data acquisition and if enough of the
images are cropped the event rate can appear lower than normal. Normally
this is due to air in the system. Run the purge bubbles script from the instru-
ment drop-down menu.
For more information, see Unstable fluidics (Air or clog in
Insufficient illu-
mination
Turn the appropriate lasers on. Set the 785 SSC laser to 40mw. Set the laser
powers to maximum and decrease them to prevent pixel saturation.
Cross-con-
tamination from
previous
samples
DNA dye from pre-
vious sample is
labeling current
sample
DNA dyes must be thoroughly flushed from the sample lines, to prevent resid-
ual dye from labeling subsequent samples. Load a sample of 10% bleach fol-
lowed by a PBS wash, to remove all traces of the DNA dye in the instrument,
or run the sterilize script (~30min).
Cells from the pre-
vious sample are
appearing in current
sample
This suggests a minor clog. Load a sample of 10% bleach followed by a PBS
wash to remove most contaminating cells, or run the sterilize script
(~30min).
Erroneous fluid
level indicator
Tank has moved away
from the sensor
Open the buffer compartment and move the tank closer to the sensor until
the fluid level indicator is correct.
Sensor is broken
Power down and power up the instrument, if this does not fix the problem,
call Luminex Technical Support.
For Research Use Only. Not for use in diagnostic procedures.
44
Amnis
®
FlowSight
®
Imaging Flow Cytometer User Manual