Start Camera: Turns on imaging if it has stopped.
l
Calibrate with ASSIST: Opens the Calibrations and Tests window.
l
Load Sheath: Fills the sheath syringe with sheath fluid and an air bubble that facilitates stable flow.
l
Load Flush Syringe: Fills the flush syringe with sheath fluid.
l
Prime: Rapidly pushes a small volume of sample into the flow cell
l
Purge Bubbles: Removes air bubbles from the flow cell by filling the flow cell with air then filling the
sheath line and pump with debubbler and rinsing the flow cell. The sheath syringe is then refilled with
sheath and the bubble trap, lines and flow cell are filled with sheath.
l
Purge Sample Load Line: Cleans out the sample load line
l
View Tank Levels: Opens the fluid level window.
l
Service Scripts: For field service personnel only.
l
Usage Monitor is available to find the information on scripts that have run
l
Options
l
Analysis menu: Access the Feature, Population and Region Managers. Functionality is the same as for IDEAS
®
Software. Refer to the IDEAS
User Manual
for more information.
l
Features: Opens the Feature Manager. Features can be renamed or new combined features can be cre-
ated.
l
Populations: Opens the Populations Manager. View, edit, or delete populations.
l
Regions: Opens the Regions Manager. View, edit or delete regions.
l
Wizards: Opens the list of Wizards including the compensation wizard.
l
Layout menu:
For Research Use Only. Not for use in diagnostic procedures.