3. Turn on each laser used in the experiment by clicking on the wavelength. Observe the scatter plots of Raw Max
Pixel for the channels used in the experiment. Set the laser powers so each fluorochrome has Raw Max Pixel Intens-
ities between 100 and 4000 counts, as measured in graphs and there is no saturation. The default saturation color
on the images is cyan but can be changed in the display settings section.
In the examples below, the objects are not saturating in channels 2, 5 or 7 but saturating in channel 11 indicated by
the events lining up at 4x10
3
counts.
4. Select Brightfield channels. Default is Channels 1 and 9. Click Set Intensity.
5. Set the display properties if needed,
see Setting the Image Display Properties
For Research Use Only. Not for use in diagnostic procedures.
20
Amnis
®
FlowSight
®
Imaging Flow Cytometer User Manual