Your IT department must provide the following information:
l
SMTP Host
l
SMTP Port (add firewall exceptions for INSPIRE and this port)
l
Login
l
Password
l
To
l
From
l
Cc (optional)
l
Choose whether to enable SSL
If the email notifications do not work, have your IT department make sure the messages are not being blocked
by firewalls or other network security settings. Also, try enabling / disabling the SSL check box.
Sample Validation:When using the AutoSampler, each sample load is validated to ensure good imagery is collected for
the sample.
l
First the core speed and rough synchrony is checked to make sure it is close enough to proceed and this
requires Brightfield (if the template has BF off then BF must be turned on). This step occurs whether validation
is ON or OFF in the plate definition.
NOTE: Luminex highly recommends that the validation is 'Yes' in the plate definition for every well.
l
When the validation is 'Yes' for the well, the object rate and synchrony is then tested and must pass some pre-
determined limits. Focus adjustment is enabled. This ensures good imagery is collected.
l
Finally, if BF was turned ON for the first step then it is turned off again for the current template.
l
If samples of low concentration will not pass the validation step, add 20 μL of Calibration beads to the samples.
For Research Use Only. Not for use in diagnostic procedures.
38