Chapter 5
_________________________________________________________________________________________
_____________________________________________________________________________
BE6020-14
1316A-1 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 97 of 98
5.1.8 O
2
and NO Sensor Replacement
The optional O
2
and NO sensors have an expected life of 18 months under
normal operating conditions. Follow the procedure defined below to replace
these sensors.
a)
Verify that main power turned OFF.
b)
Remove the cover.
2)
Sensor Removal:
a)
Each sensor includes a pull tab. Be careful not to tear it.
b)
Carefully pull on the tab and remove the sensor from the 1316A-1 Multi
Gas Monitor UA1372 manifold.
3)
Sensor Installation:
a)
Position the new sensor at the sensor manifold. Ensure that the sensor’s
connector pins are aligned.
b)
Carefully push on the sensor to seat it in the Manifold UA1372.
c)
Replace the cover.
4)
Sensor Calibration: Refer to Chapter 4.
5)
Verification: Consult the Status window and verify that no errors are
reported.
5.1.9
Field replaceable parts.
LumaSense Part number Description
DS0790
Zero ref. filter
DS2306
Inlet filter
VF0007A
Fuse 1.6A T
VM0200A
O
2
electrochemical
Sensor
VM0201A
NO electrochemical
Sensor
UA1372A
Mounting kit for
electrochemical Sensors
UH5040A
Pump
UM1118A
Tree way valve
ZG0370A
Power supply