1.
Use to tick which of the CO, CO
2
, CHx gasses that you want
to be calibrated.
2.
Type in the correct concentrations for the ticked CO, CO
2
, CHx
gas.
The gas concentration must be the same as that stated on the
calibration gas bottle.
3.
Connect the calibration-gas bottle as described in
4.
Open the gas valve and adjust the flow so that a small surplus
flow exists in the side-stream tube.
5.
Press the Send button to start the gain calibration.
After making the Span Calibration please make sure that the Mode is
set back into 5 and that the Zero calibration is set to “Auto”, as
described in
The monitor is now ready for making measurements.
Press Send.
The Process State In Progress is displayed in the Span Calibration
) and the Data Status in the Status window is
The 1316A-1 is calibrated with the selected gas at the concentration
keyed in.