Chapter 3
_________________________________________________________________________________________
_____________________________________________________________________________
BE6020-14
1316A-1 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 38 of 98
Figure 3.3. BZ6012 Main Window showing measurement data. Upper
part giving the mean value, Middle part the standard deviation value
for the selected number of samples. Bottom part displays the real
time measurement values.
To change the settings select, none, mean or mean+STDEV and the
number of samples to use during averaging. Press OK and then
START the window is now updated according to the new settings.
Zero
It is possible to select either Auto or Manual.
If Auto is selected then the Monitor will automatically perform zero
calibrations upon request from the Module. Notice that during a zero
calibration the Oxygen sensor (option) will be span calibrated with
ambient oxygen level.
If Manual is selected press the Soft Key Zero to perform a Zero
calibration.
To see if a zero calibration is requested pull down the Screens drop
down and select Status.
The Status screen will be displayed, see figure 3.4.