Chapter 4
_________________________________________________________________________________________
_____________________________________________________________________________
BE6020-14
1316A-1 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 71 of 98
Figure 4.16 Main window displaying measurement started..
Open the status window and press the “Zero”-button in the Main window, in
order to perform a Zero calibration.
The ambient air is taken from the Zero reference inlet on the rear panel during
Zero calibration. In the Status window observe that the “Proc in prog”-status is
“In Progress”, while Zero calibration. is done, see
When Zero calibration has successfully finished the “Proc in prog”-status is “Not