3.4.1
Pump In Flow.
Not included in the 1316A-1.
3.4.2 New NO Sensor
This fault indicate that the Full-scale output of the NO sensor is
≤80% of full scale value when sensor was installed. The NO sensor
must be replaced with a new NO sensor.
Please refer to section
sensor
3.4.3 New O
2
Sensor
The O
2
sensor output is detected to be below a specified limit during
Zero/O
2
Span Calibration. Replace the O
2
sensor. Please refer to
2
sensor.
3.4.4 IR signal lost
This fault indicate that the sample Cell is contaminated. The sample
cell and windows must be cleaned. Might also display CO2, CO and
HC data invalid in the Data Status section of the status window see
3.4.5 Pump Out Flow
a. Sample gas path is occluded at outlet during normal mode
operation.
Corrective action: Clear outlet occlusion. Use the pump on command
see section 5.1.3.
b. Sample gas path is occluded at the outlet during Zero/O
2
span
calibration.
Corrective action: Clear outlet occlusion. Use the pump on command
see section 5.1.3 and perform a new Zero/O
2
span calibration see
section 4.2.1.
Might also display a Zero calibration fail error in the Data status of
the status screen.
CO
2
, CO, HC, NO Zero calibration fail and O
2
span calibration fail will
be displayed.