MODEL 2350-1 Data Logger
7
2. CALIBRATION ROUTINES
SOURCE PARAMETERS AND TYPICAL VALUES FOR MODEL 2350-1 CALIBRATIONS
To insure that the proper sized sources are used in calculating the calibration constant, and
dead time of the Model 2350-1 and detectors the following calculation should be performed.
% Dead Time Loss = No. of source counts in 1 second X Dead Time (in seconds)
The following list provides some typical LO and HI cal points and dead times for various
detectors:
TYPICAL DEAD TIME CALIBRATION, AND OVERLOAD POINTS
TYPICAL DEAD TIME AND CALIBRATION CONSTANT SETTINGS
DETECTOR
DEAD TIME
CALIBRATION CONSTANT
44-2
15 - 25
µ
sec.
1.0e
10
- 0.3e
10
44-6
65 - 135
µ
sec.
7.09e
7
- 0.65e
7
44-7
250 - 270
µ
sec.
1.39e
8
- 0.1e
8
44-9
50 - 120
µ
sec.
2.0e
8
- 0.3e
8
44-38
65 - 135
µ
sec.
7.09e
7
- 0.65e
7
4 3-68 Alpha
15 - 25
µ
sec.
1
43-68 Beta
18 - 26
µ
sec.
1
4 3-37 Alpha
18 - 20
µ
sec.
1
43-37 Beta
19 - 25
µ
sec.
1
DETECTOR
LOW PO INT
HIGH POINT
OVERLOAD POINT
44-2
500
µ
R/hr
15 mR/hr
100 mR/hr
44-6
20 mR/hr
400 mR/hr
1 R/hr
44-7
5 mR/hr
100 mR/hr
N/A
44-9
6 mR/hr
200 mR/hr
1 R/hr
44-38
20 mR/hr
400 mR/hr
1 R/hr
44-116
6 mR/hr
60 mR/hr
N/A