Model 334A Operations Manual
Ludlum Measurements, Inc
49
January 2021
Standard Log Data Format
The log data is reported in comma-delimited format. Each channel is transmitted in a string
of data, with multiple strings each minute depending on the unit configuration. The text
fields are terminated with a <cr><lf> sequence. Below is an example of 1 minute of strings.
M334A_301, 68551.00, Am241/C,
0, 0.0054, 0.0000, 0.0275, 0.0030,
5.87, 6703.4, 04711
M334A_301, 68551.00, Pu239/C,
0, 0.0043, 0.0000, 0.0189, 0.0029,
5.87, 6703.4, 04805
The following header describes the fields.
Name
Time
Chan
Status
Concen
MDC
Dose
MDD
Flow
Volume
Checksum
Where,
Name
is a string which uniquely identifies the instrument (M334A_301)
Time
is the elapsed time in seconds since the filter was changed or the instrument reset
Chan
is the name of the channel (Pu239, Am241, PAEC)
Status
is an integer value as follows:
Initializing
-1
Normal
0
Standby
40
Alarm_High 18
Flow_Low 24
Flow_High 25
Fail_Det
32
Fail_Low_Batt 33
Fail_Int_Comm 34
Fail_Poor_Fit 36
Fail_Flow 37
Fail_Calib 38
Concen
is the derived airborne concentration in units of DAC (floating point)
MDC
is the minimum detectable concentration in DAC (floating point)
DAC-h
is the estimated worker dose in units of DAC-h (floating point)
MDD
is the minimum detectable dose in units of DAC-h (floating point)
Flow
is the sample flow rate in liters per minute
Volume
is the total sampled volume since filter change/reset in liters
Checksum
is a 16-bit Checksum calculated on the characters of the string up to, but not
including, the last comma. Calculation is modulo-65,536 (2-byte unsigned integer).
Each single-line record contains the log data for one channel, depending on the instrument
type and configuration. All units will include the IoI#1 information, and if the unit has a
second IoI defined, it will include the IoI#2 information. If multiple lines of data are
requested, each data record will appear on a separate line.