Model 334A Operations Manual
Ludlum Measurements, Inc
28
January 2021
Flow Tab
This page defines the high, low, and failure limits for air flow measurement in liters per
minute. It also offers a Fixed Flow setting for externally plumbed units.
Low Limit
—
This setting defines the limit in liters per minute below which the Model
334A will indicate a LO FLOW status. A low flow status can indicate a loaded or wet
filter.
Default is 6.0 LPM
High Limit
—
This setting defines the limit in liters per minute above which the Model
334A will indicate a HI FLOW status. A high flow status can indicate a missing filter,
leaking or disconnected vacuum line. To achieve a ‘missing filter limit’, remove the filter,
replace the filter holder, check the flow rate, and set just below this rate.
Default is 7.5
LPM.
Low Fail Limit
—
This setting defines the limit in liters per minute below which the
Model 334A will indicate a FLOW FAIL status, go out of service, and will not alarm. A
flow fail status can indicate a kinked or clogged vacuum line.
Default is 4.5 LPM.
Use Fixed Flow
—
When checked, this setting is used for units with external pumps
with no flow measurement capabilities.
Default is Unchecked.