![Ludlum Measurements 334A Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/ludlum-measurements/334a/334a_operation-manual_1946971033.webp)
Model 334A Operations Manual
Ludlum Measurements, Inc
24
January 2021
Logging Tab
The
Logging Interval
settings control the data logging behavior of the Model 334A. Several
types of log files are created on the Model 334A during operation. See Section 6: Retrieving
and Analyzing the Data, for more information on the log files.
Log Acute readings every
—
This setting controls the frequency in seconds that Acute
readings are written to a log file (Pu239_AcuteYYYYMMDD.txt). Each filename
created has the format
NameYYYYMMDD.txt
, where
YYYY
represents the year,
MM
represents the month, and DD represents the creation day.
Default is 15 Seconds.
Log Spectrums
—
This setting controls whether spectrum snapshots are written to a log
file (SpectrumYYYYMMDD.txt). If the setting is checked, the Log Spectrums
frequency field will be enabled.
Default is Spectrum Logging is ON.
Log Spectrums every…—
This setting defines the frequency in minutes that spectrums
are time stamped and logged.
Default is 10 Minutes.
Log data destination
This selection defines the location where Model 334A data will be saved.
My Documents folder
—
This setting saves log files in the My Documents folder.
This
is a volatile folder that will be erased in the event of a complete battery discharge
or loss of battery power.
Default is the My Documents folder.
SD Card
—
this setting causes log data to be saved to the non-volatile SD memory card
in the Model 334A.
Note: data saved to the SD Card is accessible from an attached PC/laptop, but
data is not automatically synchronized with the PC as are files in the My Documents folder.