![Ludlum Measurements 334A Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/ludlum-measurements/334a/334a_operation-manual_1946971010.webp)
Model 334A Operations Manual
Ludlum Measurements, Inc
1
January 2021
Getting Started
General Description
The Model 334A is a portable alpha air monitor designed to function both as a workplace
monitor and a Continuous Air Monitor (CAM) for measurements. Spectral analysis is conducted
via a 1024-channel analyzer that feeds data to the embedded processor. Factory configuration
provides either special nuclear materials (SNM) or radon progeny measurements of potential
alpha energy concentration (PAEC).
Measurements are taken in both fast-responding (acute) and high-sensitivity (chronic) assessments
and report in English or SI units.
The instrument stores acquired data in comma-separated-variable (.csv) format. Data may be
saved in the instrument’s internal memory or written to an SD card for later retrieval.
The Model 334A features an integrated LCD and touchscreen that displays information on
instrument status and readings during operation. The estimated dose of the isotope(s) of interest
and instrument status is displayed at all times. A window below may be switched from showing
historical readings and battery status, or to display the current spectrum.
Unpacking and Repacking
Remove the calibration certificate and place it in a secure location. Remove the instrument and
ensure that all of the items listed on the packing list are in the carton. Check individual item serial
numbers and insure calibration certificates match between instruments.
To return an instrument for repair and calibration, provide sufficient packing material to prevent
damaged during shipment. Every returned instrument must be accompanied by an Instrument
Return Form, which can be downloaded from the Ludlum website at www.ludlums.com. Find
the form by clicking the “Support” tab and selecting “Repair and Calibration” from the drop-
down menu. Then choose the appropriate Repair and Calibration division where you will find a
link to the form.