Model 334A Operations Manual
Ludlum Measurements, Inc
16
January 2021
After the calibration has completed, it will report the efficiency. Typical 4pi efficiency is 20-
30% for Am-241. If the efficiency falls outside these parameters, either
Repeat Calibration
or
call a Ludlum Technician for assistance.
If the efficiency is acceptable, remove the source from the detector, place the filter back, and
close up the detector head. Tap
OK
to close the dialog and save the new efficiency setting.
Automatic Alpha Calibration
Customers should first perform the Advanced Calibrations, using the same calibration
isotope source, before running Automatic calibrations.
The automatic alpha energy calibration is designed to make regular alpha energy calibration quick,
easy, and accurate. This calibration mode will adjust the offset parameter in the MCA and the
energy channels of the radon isotopes to compensate for a shift of the spectrum due to electronic
drift over time, or from significant temperature and pressure changes.
To start the Automatic calibration tap
Calibrate—Automatic Alpha…
on the menu. Open up the
detector head and remove the filter. Insert the source defined as the “Alpha Source” and tap
Start Calibration.
The calibration routine will calculate the alpha efficiency and display the efficiency results. The
Alpha 4-Pi Efficiency should be 20-30 %. An efficiency percent outside of the expected range
would indicate a discrepancy in the process or possibly with the unit.
Repeat Efficiency
or call a
Ludlum Technician for assistance. Hit
Next
to go to the Calibrate Energy step.