Model 334A Operations Manual
Ludlum Measurements, Inc
17
January 2021
In the Calibrate Energy step, the peak fitting process occurs. This is where the red lines that
correspond to the isotopes’ energy levels (peak channels) line up with the peaks displayed by the
spectrum. The peak should appear with the calibration source still inserted in the detector head.
If the peak channel does not line up, then the user needs to drag the red marker to the height of
the peak. Once the peak channel is lined up, select
Finish
. If you elect to
save
the new settings,
then the offset, radon peaks and second IoI peak will be adjusted an equal amount to compensate
for the new attenuation.
For most users doing a re-calibration using the same isotopes, these Automatic Alpha steps will
be sufficient to calibrate the instrument. Instruments that exhibit POOR_FIT status or erratic
readings should be recalibrated using the advanced calibration procedure.
Response Check
After a calibration has been performed, it is good to double check the calibration by running
a Response Check. The response check can also be done at any time if the user wants to
verify the unit is still in calibration. Follow the detailed instruction is Section 4 of this
manual.