General Tab
Serial No
—
This is a
non-editable
field that identifies the factory-defined serial number of
the Model 334A.
Default: Factory set
Window Settings
Acute Window
—
This setting defines the count time used to determine the Acute Dose,
NetCps and Activity. It should be set to provide the fastest response to high
concentration spikes of a pre-determined level. Shorter times mean higher minimum-
detectable levels but do not necessarily mean faster response times. Some
experimentation may be required to achieve a desired alarm level and response time.
Default is 120 Seconds.
Chronic Window
—
This setting define the count time used to determine the Chronic
Dose, Concentration, NetCps and Activity. Longer times provide better sensitivity and
lengthen the dose integration time.
Default is 240 Minutes.
Calibration Expiration Settings
Calibration expires
—
This non-editable item displays the latest date on which the
Model 334A calibration will expire. The expiration could occur sooner if the maximum
runtime hours are exceeded.
Calibrations are valid for XXX days
—
This field sets the maximum number of days
between calibrations.
Default is 365 days.
But not more than YYY runtime hours
—
This field sets the maximum number of
hours of Model 334A runtime between calibrations.
Default is 8800 hours.