Model 2241-2RK
Technical Manual
Section 8
Ludlum Measurements, Inc.
Page 8-15
July 2012
CC and DTC Algorithm
Equations (5) and (6) convert units per time (R/hr Display Units) to
units per second:
uni t s
t i me
uni t s
s ec ond
⇒
Insert the cps
lo
data point (8 mR/hr for the Model 44-9 example)
determined from equations (1) and (2):
Insert the cps hi data point (200 mR/hr for the Model 44-9 example)
determined from equations (3) and (4):
Place the detector in the low field and enter the counts per second:
Note:
The low field count sample should be
≥
3000 counts. Use the
Scaler and adjust the count time to accumulate count
≥
3000.
Equation 5
s
10
x
2.22
=
)
s
60
m
1
(
x
)
m
60
h
1
(
x
)
h
R
(0.008
=
CAL
6
lo
−
Equation 6
s
10
x
55.6
=
)
s
60
m
1
(
)x
m
60
h
1
(
)x
h
R
(0.200
=
CAL
hi
6
−
Equation 7
s
counts
=
time
count
SAMPL
=
CORR
lo
lo
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......