![Ludlum Measurements 2241-2 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/ludlum-measurements/2241-2/2241-2_manual_1946943014.webp)
Model 2241-2RK
Technical Manual
Section 3
Ludlum Measurements, Inc.
Page 3-1
July 2012
Specifications
0
Linearity
: Readings are within 10% of true value with a detector
connected.
Warm-up Time
: Unit may be used immediately after the LCD
initialization sequence is completed (approximately five seconds after
power-up).
Display
: a four-digit Liquid Crystal Display (LCD) with digits one half
inch in height. Two additional 0.51 cm (0.2 in.) digits are used for the
overflow counter (
SCALER
mode) and exponential powers (parameter
setup). Enunciators are provided for display units,
ALERT
,
ALARM
, low
battery, detector
OVERLOAD
, counting
OVERFLOW
, and scaler
COUNTING
.
RATEMETER
: depending upon how the instrument was calibrated, the
RATEMETER
can display in R/hr, Sv/h, cpm, or cps when the control
switch is in the
RATEMETER
position.
SCALER
: activated by pushbutton in handle when the three-position
switch is in the
SCALER
position. Count time is adjustable.
Calibration Controls
: accessible from the front of the instrument
(protective cover provided). These controls are preset at the factory or
calibration lab and should not be adjusted by field personnel.
Discriminator / Input Sensitivity
: adjustable from 2 to 100
mV; negative pulse response
Overload
: indicated by
OVERLOAD
on the display; adjust-
able depending on the detector selected
High Voltage
: two independent controls, each adjustable
from 200-2500 volts, two separate set points for
DET
1 and
DET
2, regulated within 0.2% at 1000 Vdc; maximum load of
50
µ
A
RESET
: a pushbutton for zeroing the display, acknowledging and/or
resetting the alarm
Section
3
Instrument
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......