Model 2241-2RK
Technical Manual
Section 2
Ludlum Measurements, Inc.
Page 2-3
July 2012
The display will auto-range to the current level (see figure at left).
When auto-ranging down, the Model 2241-2 uses multiples of 5.
This technique keeps the decimal point from jumping between
numbers when viewing values around multiples of 10.
Check for a proper background reading:
If using a Ludlum Model 44-9 detector, a typical reading
would be 25-50 cpm or 8-15
µ
R/hr.
If using a Ludlum Model 44-2 detector, a typical reading
would be 1.4-2.6 kcpm or 8-15
µ
R/hr.
A reference reading (or readings) with a check source should be obtained
with the detector(s) in a constant and reproducible manner at the time of
calibration or at the time the instrument is received in the field.
If at any time the instrument fails to read within 20% of the reference
reading when using the same check source, it should be sent to a calibration
facility for recalibration and/or repair. If desired, multiple readings may be
taken at different distances and/or with different sources so that other
ranges or scales are checked.
Switch the
AUD ON
/
OFF
switch to the
ON
position and confirm that the
external unimorph speaker produces an audible click for each event detected
(audio divide-by 1 parameter). The
AUD ON
/
OFF
switch will silence the
clicks if in the
OFF
position, however, an audible alarm condition will still be
heard.
Increase the source activity or lower the alert and alarm points to initiate an
ALERT
and
ALARM
condition. Refer to Section 8, “Entering or Changing
Switchboard Parameters.” Depress the
RESET
switch to acknowledge the
audible alarm. Decrease the radiation activity below the
ALERT
and
ALARM
threshold and depress the
RESET
switch to clear the alarm conditions. If an
alarm condition is not present, depressing the
RESET
switch the first time will
reset the alert condition and zero the ratemeter.
Position a check source to produce a ratemeter reading of 100 to 2000
counts per minute or 10-100 µR/hr. While observing the ratemeter
fluctuations, select between the fast and slow response time (
F
/
S
) positions
to observe variations in the display. The
S
position should respond
approximately five times slower than the
F
position (for fixed response
mode) and three times slower when in variable response mode. The slow
response position is normally used when the Model 2241-2 is displaying low
numbers that require a more stable display. The fast response position is
used at high count levels.
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......