![Ludlum Measurements 2241-2 Скачать руководство пользователя страница 55](http://html.mh-extra.com/html/ludlum-measurements/2241-2/2241-2_manual_1946943055.webp)
Model 2241-2RK
Technical Manual
Section 8
Ludlum Measurements, Inc.
Page 8-14
July 2012
Reference the table at the end of this section to determine the
cps/exposure rate (cps/ER). The conversion can be determined by
placing the detector in a radiation field
which produces from 50 to 200 cps.
Calculate the count/exposure rate
using the equation to the left.
For example, exposing a LMI Model
44-9 to a 2 mR/hr
137
Cs field yields approximately 110 cps so that:
cps/mR/hr
55
=
mR/hr
2
cps
110
The typical dead time for a M44-9 is approximately 85 µs. Therefore,
using 85 µs for “DT” in equations 1-4, the
lo
field should be between
240 and 619 cps, and the
hi
field is between 5040 and 17,650 cps.
Dividing the cps values by the 55 cps/mR/hr conversion equates to
between 4-11 mR/hr for the
lo
field and 91-320 mR/hr for the
hi
field.
Select a calibrated field between the
lo
and
hi
data points determined
above:
lo
(CAL
lo
) = 8 mR/hr
hi
(CAL
hi
) = 200 mR/hr
The following procedure outlines the hi-lo method
units = Sv, R, counts.
CAL
lo
=
lo
field calibration point.
CAL
hi
=
hi
field calibration point.
CORR
lo
= recorded field at low calibration point.
CORR
hi
= recorded field at high calibration point.
DT = dead time constant entered into Model 2241.
CC = calibration constant entered into Model 2241.
f
d
and
a
d
are intermediate steps in calculating DT.
f
cal
is an intermediate step in calculating CC.
cps
radiation field in exposure rate units
= cps / ER
Abbreviations used:
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......