mode when the six-digit display (four digits display and two overflow
digits in right corner) reaches 999999 and starts to roll over again.
OVERLOAD
: This indicates that the detector is being exposed to
radiation intensities greater than the detector maximum operating limit.
For alpha and/or beta-type scintillation detectors, an
OVERLOAD
may
indicate that the detector face has been punctured, allowing external
light to saturate the photomultiplier tube inside the detector. The
overload alarm point is set by adjusting the
OVL
control located
underneath the calibration cover.
Note:
Overload is set for detector Model 44-9 only.
Low battery icon
: This indicates that the batteries have decreased to
the minimum operating voltage of 2.2 ±0.1 Vdc. Instrument will
continue to operate for approximately 24 hours thereafter.
COUNTING
: This indicates that the scaler
COUNT
switch has been
depressed and that the scaler is accumulating counts for the pre-
determined count time.
Front Panel Controls
DET1/DET2 (Detector Select) Switch
: a protected (locking) toggle
switch used to select between
DET
1 and
DET
2 parameters. The switch
can be pulled up and then over to the desired selection.
OFF/RATEMETER/SCALER Switch
: a three-position rotary switch that
applies power to the instrument and selects
RATEMETER
or
SCALER
counting mode.
AUD ON/OFF Switch
: The clicks-per-event audio may be silenced or
enabled via this front-panel toggle switch. The audible alarm is
independent of the
AUD ON
/
OFF
switch and will override the audible
clicks-per-event. An audible alarm can only be silenced by depressing the
RESET
button.
F/S (Fast/Slow) Response Switch
: a two-position toggle switch that
selects fast or slow counting response time
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......