![Ludlum Measurements 2241-2 Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/ludlum-measurements/2241-2/2241-2_manual_1946943013.webp)
Model 2241-2RK
Technical Manual
Section 2
Ludlum Measurements, Inc.
Page 2-5
July 2012
Compare the first reading with the second to determine the
amount of alpha and beta contribution. (This will be a very
rough number.)
Gamma and X-ray
: For gamma and high energy X-ray, the gamma
scintillator (Model 44-2) must be used.
This detector is approximately 50 times more sensitive than
the pancake (Model 44-9) detector. This detector is able to
detect low-level gamma and X-rays in the range of 60 keV-
1.25 MeV. This detector may be used for measurements up
to 50 mR/hr.
The Model 44-2 is used by placing the detector in hand, and
by slowly moving it from side to side over the area to be
measured.
Be aware that the very end of the detector, opposite the
connector, is the most sensitive. Point the detector towards
the radiation being measured.
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......