Model 2241-2RK
Technical Manual
Section 4
Ludlum Measurements, Inc.
Page 4-1
July 2012
Identification of Controls and
Functions
Display
The Model 2241-2 utilizes a four-digit liquid crystal display (LCD) with a
two-digit overflow (
SCALER
mode) and moving decimal point. The two
smaller digits located in the lower right corner of the display indicate counter
OVERFLOW
when in the scaler counting mode (equivalent to a six-
digit scaler) or exponential power when in the parameter setup mode.
The upper right corner of the LCD displays units and multiplier(s) -
R/hr, mR/hr, or µR/hr; Sv/h, mSv/h or µSv/h; C/m, kC/m, C/s or
kC/s. The bottom part of the readout displays the
ALARM
,
ALERT
,
OFLOW
,
OVERLOAD
annunciators and the low-battery icon.
COUNTING
indicates that the scaler mode has been initiated and is in the
counting process.
Display Status Definitions
ALARM
: Ratemeter or scaler count has increased above the preset alarm
threshold. An audible continuous tone will accompany the
˝
latching
˝
ALARM
condition. Depressing
RESET
will acknowledge the audible
ratemeter and/or scaler alarm. Depressing
RESET
a second time will
reset the ratemeter reading and ratemeter alarm. To reset the scaler
ALARM
, depress the
COUNT
switch located in the carrying handle to re-
initiate the scaler count cycle.
ALERT
: Ratemeter count has increased above the preset alert threshold.
To reset an
ALERT
condition, press
RESET
once if in the non-alarm
condition and twice if in an alarm condition. (The first depression in the
alarm condition acknowledges the audible alarm.) The ratemeter will
reset to the minimum displayable reading each time the alert is reset.
OFLOW
(Overflow)
: In
RATEMETER
mode
it
indicates that the incoming
count exceeds the capability to display stable or reliable readings
corresponding to the radiation level being measured. The overflow
symbol will appear when the ratemeter exceeds 100 kcps or if the dead
time correction is greater than 75%.
OFLOW
will appear in the
SCALER
Section
4
Содержание 2241-2
Страница 1: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 2241 2RK RESPONSE KIT July 2012 Serial Number 237627 and Succeeding Serial Numbers ...
Страница 3: ......
Страница 62: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 July 2012 ...
Страница 63: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 22 July 2012 ...
Страница 64: ...Model2241 2RK TechnicalManual Section8 Ludlum Measurements Inc Page 8 23 July 2012 ...
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ...Gamma Energy Response for Ludlum Model 44 9 ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......