10 Calibration
28
10 Calibration
The HPT63 provides the facility within the
Calibration
menu to adjust the sensor
inputs and current output levels to tailor the unit to the system in which it is operating.
The menu structure is as follows:
Calibration
Mode:
| On-line
| Off-line
Cal Access Y/N
Sensor Buffer
|
Are You Sure?
|
Buffer value displayed/adjusted
|
Buffer value displayed
Sensor Slope
|
Are You Sure?
|
Slope value displayed/adjusted
|
Slope value displayed
Temperature Calibration(if TC mode set to Auto)
|
Are you sure?
| Temperature
reading
displayed/adjusted
| Offset
displayed
Current Output A/B calibration
|
Are you sure?
|
Adjust 0 mA
|
Adjust 4 mA
|
Adjust 20 mA
Reset User Cal
|
Are you sure?
| Resetting
10.1 On-Line/Off-Line
Operation
Selecting the
Mode
menu will allow the user to place the unit in the
Off-line
state. If
the state is set to
Off-line
the relays will be de-energised and the current output level
frozen for the duration of the
Off-Line
state. When
On-Line
is selected the relays
and current output will operate normally. The middle LED on the front panel display
will indicate when the unit is
Off-Line
.
10.2 Calibration
Access
A feature has been included in the HPT63 to allow users to access the sensor
calibration from the main display. To enable this feature, set
Cal
Access
?
to
Y
. To
disable this feature, thus preventing the users from altering the calibration from the
main display, set
Cal Access
to
N
.