5 Main Display
17
5 Main
Display
There are two modes of display operation – digital and bargraph. The mode can only
be changed via the Configuration menu, requiring a Level 2 access code.
In bargraph mode, the display shows an analogue representation of Output A or B, or
both, alternating. The output identity and units are shown in the upper display.
In digital mode, the display shows the sensor reading, in the units assigned via
Parameters/Units
, on the top row and a secondary reading on the bottom row. The
user can select the desired secondary reading parameter by means of the
and
keys. If bargraph display has been enabled, digital readings can be
selected but the bargraph display will be restored after a timeout period. The
secondary parameters vary according to the instrument configuration, and are as
follows:
Sec. display Comment
Temp. input
Units C or F, as selected via
Parameters/Temperature Units
Output A
This shows the output current from Channel A in mA.
Output B
This shows the output current from Channel B in mA.
SP1
This appears when Set Point 1 is enabled and trigger is set to Low or High.
It shows set point value in the assigned units.
SP1H
This appears when Set Point 1 is enabled and trigger is set to Band, Latch Lo
or Latch Hi. It shows set point upper value in the assigned units.
SP1L
This appears when Set Point 1 is enabled and trigger is set to Band, Latch Lo
or Latch Hi. It shows set point lower value in the assigned units.
SP2
This appears when Set Point 2 is enabled and trigger is set to Low or High.
It shows set point value in the assigned units.
SP2H
This appears when Set Point 2 is enabled and trigger is set to Band, Latch Lo
or Latch Hi. It shows set point upper value in the assigned units.
SP2L
This appears when Set Point 2 is enabled and trigger is set to Band, Latch Lo
or Latch Hi. It shows set point lower value in the assigned units.
The default secondary parameter can be set by selecting the parameter and pressing
the
key (provided that no error messages are present). This is the parameter
that is displayed on the bottom line when the unit is switched on, or as a result of a
timeout back to the normal display mode.
If the unit has detected an error, an error message will flash alternately with the
secondary parameter or bargraph. See
Appendix F – Error Messages
.
Содержание HPT63
Страница 1: ...HPT63 pH REDOX TRANSMITTER OPERATION GUIDE...
Страница 2: ......
Страница 8: ...Contents iv...
Страница 14: ...3 Installation 6 63mm 59 mm 138mm 116mm 273 mm 70mm 116mm Figure 2 Head mounting component dimensions...
Страница 52: ...15 Guarantee and Service 44...
Страница 58: ...18 Appendix C Redox Solutions 50...