23. Adjust the X-Y stage of the APD to maximize the voltage at the “analog” BNC
on the interferometer base. This voltage is negative. Adjust the stage to
obtain smallest absolute value. A digital voltmeter or oscilloscope is
recommended for this procedure. In both the X and Y directions, there
should be a “plateau” of stable, less negative voltage. Adjust both the “X”-
screw (accessed through a hole on the rear of the I.B.) and “Y”-screw
(accessed by “grabbing” the knurled knob of the XY stage) so that the APD is
in the center of both “plateaus”.
24. Replace the rear plate on the interferometer base and tighten all screws.
4-12
Содержание Micro-g LaCoste FG5
Страница 7: ...Table of Figures This Page is Intentionally Blank vii ...
Страница 8: ......
Страница 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Страница 21: ...Design Components and Function 2 Figure 2 8 Side view of interferometer optics and beam path 2 11 ...
Страница 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Страница 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Страница 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Страница 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Страница 56: ......
Страница 59: ...Adjustment and Maintenance 4 4 3 ...
Страница 78: ...Adjustment and Maintenance 4 4 22 ...
Страница 106: ......
Страница 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Страница 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Страница 144: ......