Adjustment and Maintenance
4
Reverse the procedure for reassembly.
4.1.1.6.Replacing the V-Plate
The V-plate contains three tungsten V's which support the test mass. Since
removal and insertion of the tungsten V's in the V-plate requires special tools,
this cannot be done in the field. However, an entire new V-plate assembly can be
installed. The dropping chamber must be vented and opened for this procedure
(see “Removing the Dropping Chamber Cover” procedures above).
Remove the bottom drag-free cover
from the cart by removing the three M3
screws. Detach the LED bracket from the side of the cart and pull it out of the
way, being careful not to detach or damage the wires connected to it. Remove
the two M3 screws and the threaded post attaching the top drag-free cover to the
cart, observing the position of the post. Gently lift the cover off, being careful not
to damage the wires connected to it.
To remove the V-plate, the test mass must be partially disassembled. First, note
the orientation of the top part of the test mass and the V-plate. It is very
important to replace these parts in the same orientation. Remove the three
beryllium copper M3 screws that secure the top part of the test mass to the three
posts which pass through the V-plate. Now the lower portion of the mass can be
lowered through the V-plate. Remove the six M2 screws holding the V-plate to
the cart.
Reverse the procedure to replace the plate.
The tungsten balls from which the test mass is held are part of the top hat
assembly. If these balls need replacement, send the entire mass assembly to
Micro-g Solutions for installation and rebalancing. Reassemble the test mass
using the three beryllium copper M3 screws, and pack it carefully before
shipping. All the pieces must be included for the balancing to be done correctly.
4-6
Содержание Micro-g LaCoste FG5
Страница 7: ...Table of Figures This Page is Intentionally Blank vii ...
Страница 8: ......
Страница 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Страница 21: ...Design Components and Function 2 Figure 2 8 Side view of interferometer optics and beam path 2 11 ...
Страница 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Страница 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Страница 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Страница 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Страница 56: ......
Страница 59: ...Adjustment and Maintenance 4 4 3 ...
Страница 78: ...Adjustment and Maintenance 4 4 22 ...
Страница 106: ......
Страница 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Страница 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Страница 144: ......