How to Set Up and Run the FG5
3
8. Begin observations by executing Process | Go with the “g” program. Consult
the FG5 Software Manual for more information.
3.3. Shutting Down the FG5
3.3.1.COMPUTER
1. Exit “g”.
2. Backup the data if desired.
3. Shutdown the computer.
3.3.2.SUPERSPRING
3. Turn the COIL switch off..
4. Engage the superspring travel lock. The arrow on the lock know points
down
when it is
locked
and
up
when it is
unlocked
.
5. Disconnect the cable from superspring electronics.
3-18