9.4.8.Replacing Shaft Bearings (Drive
Pulley)
Replacing Shaft Bearings (Drive Pulley) Checklist
Remove dropping chamber cover.
Remove drive belt.
Disconnect Helical coupling between pulley shaft and Ferrofluidic
feedthrough.
Remove screws holding bottom rod ring to bottom flange.
Rotate guide rod structure so shaft clears service ring.
Remove guide rod structure.
Remove pulley yoke from bottom rod ring.
Remove bowed retaining ring from short end of pulley shaft (note
orientation).
Remove retaining ring from other end of shaft (don’t lose Woodruff
Key).
Remove pulley from bearing mounting yoke.
Push bearings out of yoke.
9-13
Содержание Micro-g LaCoste FG5
Страница 7: ...Table of Figures This Page is Intentionally Blank vii ...
Страница 8: ......
Страница 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Страница 21: ...Design Components and Function 2 Figure 2 8 Side view of interferometer optics and beam path 2 11 ...
Страница 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Страница 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Страница 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Страница 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Страница 56: ......
Страница 59: ...Adjustment and Maintenance 4 4 3 ...
Страница 78: ...Adjustment and Maintenance 4 4 22 ...
Страница 106: ......
Страница 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Страница 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Страница 144: ......