G-Link
®
-200
User Manual
56
Parameter
Specifications
Enclosure material
300 series stainless steel with polycarbonate cover
Conformal coating
TBD
Integration
Compatible gateways
All WSDA base stations and gateways
Software
SensorCloud, SensorConnect, Windows 7, 8 & 10 compatible
Software development kit
(SDK)
http://www.microstrain.com/software/mscl
Regulatory compliance
FCC (USA), IC (Canada), CE (European Union), JET (Japan)
*Actual range varies with conditions
**Measured with antennas elevated, no obstructions, no RF interferers
C Mechanical Shock Limits
The accelerometers used by the G link 200 are sensitive instruments that can suffer mechanical damage if
subjected to large or frequent accelerations beyond their rated range while in operation. To protect
themselves from occasional over-range events, the accelerometer’s control circuitry will temporarily reduce
drive power to the sense element if acceleration exceeds twice the full-scale range. While this protective
mode is enabled, the reported acceleration values will be inaccurate, and will appear to drift towards zero.
Normal operation will resume after sensed acceleration returns to an acceptable level.
It is essential to note that, even with this automatic protection circuitry, repeated exposure of the
accelerometer to shock and vibration acceleration levels exceeding twice its full-scale range will cause
excessive mechanical wear. Over time, damage can accumulate and may cause the device to become
defective. Care should be taken not to subject the sensor to accelerations beyond twice the full-scale range
during normal operating conditions
.
Содержание G-Link-200
Страница 1: ...G Link 200 Wireless Accelerometer Node LORD USER MANUAL...
Страница 2: ......
Страница 18: ...G Link 200 User Manual 18 Figure 10 Move Node...
Страница 44: ...G Link 200 User Manual 44 Figure 35 Installing The Magnetic Base...
Страница 47: ...G Link 200 User Manual 47 8 Troubleshooting a Troubleshooting Guide...