20
a.
Calibration Configuration
Each node is calibrated at the factory and stores its linear calibration coefficients in nonvolatile memory.
These coefficients are applied by the node when the node is outputting data in “floating point” format.
The coefficients are applied automatically by SensorConnect when the node is outputting data in
signed 16- bit or signed 24- bit formats. For the available data types, the calibration coefficients are
applied as follows:
l
Float (4 Bytes): Calibration applied by node
l
Int24 (3 Bytes): G-force = value * slope + offset
l
Int16 (2 Bytes): G-force = 64 * value * slope + offset
Figure 12 - Wireless Sensor Calibration Menu