The intervals of the measurements should not be too short. Intervals of 2 minutes and more
result in long pauses between the stenoses, so that venous congestion and artiospasm can be
avoided.
D.3
Abort codes BP system
Prior.
Error message
Display
Possible error cause
1
AAAA
CHECK ECG
SIGNAL
@@@@
0 / 0 mmHg
!
!
!
!
ECG trigger not connected.
!
!
!
!
ECG trigger properties not correct.
!
ECG trigger polarity not correct.
!
!
!
!
QRS pulse width
smaller than 10 ms.
2
AAAA
CUFF
OVERPRESSURE
@@@@
0 / 0 mmHg
!
!
!
!
Strong arm movements during the blood
pressure measurement.
!
Air hose is bent or blocked.
3
AAAA
AIR LEAK
@@@@
0 / 0 mmHg
!
!
!
!
Cuff is loose or not connected.
!
Air hose or air-bubble leak.
!
!
!
!
Air hose not connected.
4
AAAA
SERVICE
REQUIRED
@@@@
BP
value
!
!
!
!
Fatal error ...Get in touch with the company
Lode.
5
AAAA
CHECK ECG/
MICROPHONE
@@@@
BP
value
!
Poor patient hook-up (irregular ECG-signal) or
poor trigger signal.
AND
!
Abnormal microphone-signal
or damaged
microphone.
6
AAAA
CHECK
MICROPHONE
@@@@
BP
value
!
!
!
!
Wrong Microphone placement or defect
microphone.
7
AAAA
CHECK ECG/
ARTIFACT
@@@@
BP
value
!
!
!
!
ECG trigger signal failed during blood pressure
measurement.
Bad or irregular ECG-signal during the blood
pressure measurement.
AND
!
Excessive movements of the patient during the
blood pressure measurement.
8
AAAA
ARTIFACT/ARM
MOVEMENT
@@@@
BP
value
!
Excessive movements of the patient during the
blood pressure measurement.
9
AAAA
CUFF INFLATION
TOO LOW
@@@@
BP
value
!
!
!
!
Maximum cuff pressure adjusted too low.
!
Inflation pressure
adjusted too low. Repeat
measurement.
10
None
BP
value
D.4
Cleaning and disinfection