background image

Operator manual 

Corival 

 

 

 

30 

 

 

 
 
 

 

Caution: 

Never place the cuff so that the microphone is located in the elbow.

 

Pull the cuff so tight, that it neither moves nor ties off the arm of the patient. By fasten the 
cuff too tight or too loose, error measurements are caused. 
Wrap the end of the cuff around the upper arm and close the Velcro fastening (Fig 4: 
Mounting the cuff). 
 

 

 

 
 
 

 

Ensure that the air hose from the monitor to the cuff is not compressed, crimped or damaged. 
During ergometry the air hose should be handled and attached in such a way, that vibrations 
and movement artefacts are avoided. Use the ring strap to fix the tubes to the wrist to avoid 
tensions on the cuff due to moving and pulling tubes.

 

 

During the measurement, which starts with inflation, the arm with the cuff should be 
kept in rest. 

With ergometry it is most convenient for the patient to take the arm from the 

handlebar of the ergometer and relax the arm with the cuff. 
 

Fig 3: Cuff position 

Fig 4: Mounting the cuff 

Содержание Corival V3

Страница 1: ...Operator Manual Corival V3 English Lode BV Groningen The Netherlands 0344...

Страница 2: ...C The ergometer has to be operated under the supervision of well trained medical specialists like cardiologists pulmonologists and physiotherapists The ultimate judgment whether a test subject should...

Страница 3: ...mmon sense stated above may not eliminate all hazards Possible residual hazards could be wrong installation wrong use wrong dose wrong interpretation of readings mechanical breakdown of parts software...

Страница 4: ...Analog 12 3 4 5 TC Bytes 12 3 4 6 RS232 Protocol 12 3 4 7 Language 12 3 4 8 Device No 12 4 SPECIFICATIONS 13 4 1 TECHNICAL SPECIFICATIONS 13 4 2 FUSES 13 5 COMMUNICATION PORTS 14 5 1 RS232 14 5 2 LOO...

Страница 5: ...cuff 29 D 2 2 Position of the cuff 29 D 3 Abort codes BP system 31 D 4 Cleaning and disinfection 31 D 5 Cleaning of the cuff 32 D 6 Installing the BP cuff 33 E SpO2 option 928840 34 E 1 Precautions fo...

Страница 6: ...st subject 3 Down button 4 Up button 5 Confirmation button 6 Ergometer type indication The back panel of the Corival contains the following elements 1 Mains on off switch 2 Mains connector 3 Fuse 4 Fu...

Страница 7: ...et correct 3 Connect the power cord first to the socket of the ergometer and then connect it to the line voltage 4 Switch the power switch on The indicator lamp in the switch lights up If the initiali...

Страница 8: ...the basic Corival 1 Connect the Corival to the external control device i e ECG system PC using a correct RS232 cable 2 Switch on the ergometer and select the menu TERMINAL on the display 3 Be sure tha...

Страница 9: ...herever you are in the menu The complete menu structure is shown below In front of each menu the level depth is indicated 0 MAIN MENU 1 ANALOG 1 MANUAL 1 TERMINAL 1 SYSTEM PARAMETER 2 SET DISPLAY 3 PO...

Страница 10: ...h ON or OFF the Cadence readout in Revolutions per minute The rpm can always be read out by the LED bar graph on top of the display independent of this adjustment 3 1 3 Heartrate Switch ON or OFF the...

Страница 11: ...hat this fixed torque mode is not the same as the constant Torque mode which is used with the Wingate software module see for more information the Wingate manual 3 3 5 Isokinetic only available for Ex...

Страница 12: ...ions You need a safety code to enter the TESTFUNCTIONS 3 4 4 P In Out Analog The ergometer can be controlled by an external device which has an analog output only This is done using the ANALOG mode Wi...

Страница 13: ...resolution 1 W Read out pedalling speed analog 0 180 rpm or digital 0 255 Output workload 100 200 or 500 W 1 V max 999 W Output pedalling speed 100 min 1 1 V Input voltage for external control 100 20...

Страница 14: ...ontrolling device and the Pin layout RS232 connector 5 2 Loop connection It is possible to control several Corival ergometers with one external device using RS232 Connect the RS232 in connector to the...

Страница 15: ...eaned with a cloth dampened with non aggressive liquid A 2 Replacing fuses You will not need a certified technician to replace fuses of the Corival Always disconnect the power cord first Unplug the ma...

Страница 16: ...ometer B 2 Control panel Corival with Programmable Display Unit option 928810 1 Display B 4x20 characters directed to operator 2 Infrared window only active for option 928820 3 One step back button 4...

Страница 17: ...enu appears The menu displayed in capitals can be selected As extra reminder the first 4 letters of the selectable mode are showed above the confirmation button With the keys you can scroll through th...

Страница 18: ...ur Sport 925900 1 PROTOCOL 2 PROTOCOL EDIT 2 PROTOCOL RUN 3 STRAND PROTOCOL 2 PROTOCOL DELETE Compared to the standard version the menus are added B 3 2 Manual menu Selecting the manual mode will show...

Страница 19: ...if HR option is being installed The ergometer is now ready for digital external control For more information about digital external control using the RS 232 serial port see the user guide supplied wit...

Страница 20: ...e menu SETTINGS will give you the opportunity to select PROTOCOL as a default startmenu besides the previously described menus ANALOG MANUAL TERMINAL and SYSTEM PARAMETERS An extra option for the Prog...

Страница 21: ...en made and the menu NEW With the keys you can select the desired protocol in this example name1 to edit Select NEW to program a new exercise protocol and give the protocol a useful name maximal 10 ch...

Страница 22: ...After the confirmation of the workload during the recovery stage the menu PROGRAM BPM will appear Toggle with the keys for the measurement possibilities No BPM no BPM module available select this opti...

Страница 23: ...P with the keys and change the Workload to 150 in duration of 300 seconds press the confirmation button 12 Select P with the keys and change the Workload to minus 60 in 60 seconds press the confirmati...

Страница 24: ...appear on screen To scroll through the total list use the arrows for instance If you want to select e g the protocol 10 w min use the key until a space appears in front of the name of the protocol the...

Страница 25: ...appendix concerning Blood pressure in this manual The next steps will automatically follow With the buttons you can go the next stage or previous stage when you do not want to follow the protocol as p...

Страница 26: ...till be measured when BPM is displayed at the confirmation button To stop the recovery phase press the ESC button to go back to the PROTOCOL menu In the menus MANUAL TERMINAL or ANALOG the BP function...

Страница 27: ...ck following tips Wet the contacts on the right and left backside of the Polar belt for good contact Stay near the receiver within a range of 0 8 meter which is located in the display unit at the test...

Страница 28: ...ing menu appears In the upper line of the display still your selected readout parameters appear The buttons below have the following functions dis Toggle the display between your readout parameters an...

Страница 29: ...important Experience has shown that a lot of troubles can be avoided when the rules below are kept in mind Especially BP measurement using Korotkoff tones under stress requires accurate preparation D...

Страница 30: ...from the monitor to the cuff is not compressed crimped or damaged During ergometry the air hose should be handled and attached in such a way that vibrations and movement artefacts are avoided Use the...

Страница 31: ...not connected 4 A A A ASERVICE REQUIRED BP value Fatal error Get in touch with the company Lode 5 A A A ACHECK ECG MICROPHONE BP value Poor patient hook up irregular ECG signal or poor trigger signal...

Страница 32: ...ir hose removal Now you can open the Velcro fastening of the cuff and take out the microphone Fig 6 Microphone removal The cuff can be washed at 30 C The air hose can be cleaned with mild non alcoholi...

Страница 33: ...the Velcro fastening D 6 Installing the BP cuff To install a BP option purchased afterwards follow the instructions given in the user guide supplied with the ergometer Installation of the BP module it...

Страница 34: ...he User Guide for your Corival Contraindications Warnings and Cautions are stated The user should have knowledge of this information E 2 Using the SpO2 Option The SpO2 comes together with the P Displa...

Страница 35: ...pulse signals MIDA SpO2 could not be calculated missing data indicator ARTF Artifact A detected pulse beat didn t match the current pulse interval MPRF Marginal Perfusion Amplitude representation of m...

Страница 36: ...get rpm is shown The default value is 50 The current rpm is shown on the upper right The maximum target rpm is 80 the minimum value is 30 rpm With the up and down key s the target rpm can be changed t...

Страница 37: ...the left The current rpm is shown on the right With the up and down key s the target rpm can be changed to the desired value On the first button press the 0 Watt startup system starts when the curren...

Страница 38: ...S232 Protocol Lode Ergometer Agilent HP StressWriter M2488A Lode WLP Bosch EKG 506 DS Bosch EKG506DS BTL BTL 08PC LODE Protocol BTL 08 Win Ergo software LODE Protocol Burdick Spacelabs Spacelab LODE P...

Страница 39: ...ol Beijing Meigaoyi Co LTD ECG LAB Treadmill Stress System Lode WLP Mortara X scribe II LODE Protocol Neurosoft Poly Spectrum Velo P10 Nihon Kohden Cardiofax V ECG 93xx series P10 Norav PC ECG 1200 LO...

Страница 40: ...the RS232 protocol by following the procedure below Switch on the ergometer and press enter You will be in the MAINMENU Select by using the up down keys the menu SYSTEM PARAMETER and press enter Sele...

Отзывы: